PART B : 6 Installing or Removing the Bellows
3
Remove the bellows from the vertical shaft.
1 Turn the up/down handle counterclock-
wise again to lower the specimen stage.
2 Pullthebellowsupandothevertical
shaft.
Tips
A guide is attached to the bottom of
the bellows. Be careful not to lose it
when removing the bellows.
②
①
③
No. Name
①
Vertical shaft
②
Bellows
③
Guide
After you remove the bellows, be sure to put the inner rings (x2) on the vertical shaft.
Withouttheinnerrings,yourhandsorngerscouldgetpinchedbetweentheanvilandtheuppersurfaceofthe
vertical shaft when the specimen stage is lowered to the bottom edge.