318 Installation Testing and Verification
Calibration Offsets
The calibration offsets graphical report shows the results for tests of tape magazine, drive sled, and I/E
station offsets compared to predefined tolerances. Reports are generated for drive-side and door-side for
all frames.
• If the graphical report shows a failed result for one or more tape magazines, make sure the
magazines at the location of the failure are installed properly and that the calibration targets are
correctly snapped on to the magazines.
• Make certain the magazines are fully seated in the I/E station.
• If the graphical report shows a failed result for the I/E station, make sure the I/E station and front
door are completely shut.
• If running the test again still generates failed results, realign the middle X-axis rail and check the
alignment of the top and bottom X-axis rails at the location of the failure.
Figure 30 Calibration Offsets Graphical Report