ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-5
No. Message Interpretation Operator action
165 LIS/HIS: High-
level protocol
could not
generate high-
level packet
An error occurred while
formatting a message.
- Check protocol
configurations.
Contact Radiometer
service representative.
166 LIS/HIS: General
communication
error
An internal error occurred in
the LIS/HIS communication
module.
- Contact Radiometer
service representative if
the problem persists.
167 LIS/HIS: High-
level protocol
received packet
in wrong format
An error occurred while
parsing (interpreting) a
message.
- Check protocol
configurations.
Contact Radiometer
service representative.
200 User msg: This is only a message. An
operator has entered a note
in the log.
- No action required.
201 Westgard Rule
(1.2s) violation
Measured parameter value is
outside the mean +/- 2 SD
range.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
202 Westgard Rule
(1.3s) violation
Measured parameter value is
outside the mean +/- 3 SD
range.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements including
electrodes.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
203 Westgard Rule
(2.2s) violation
Two consecutive
measurements are outside
the mean +/- 2 SD range on
the same side of the mean.
This may indicate a shift.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements including
electrodes.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
204 Westgard Rule
(R.4s) violation
The difference between two
consecutive measurements
exceeds 4 SD. This may
indicate an inconsistency in
your procedure or an
unstable analyzer.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
electrode replacements.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals