10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-36
No. Message Interpretation Operator action
1195 ABL not
connected to QA
Portal - incorrect
password
The analyzer was refused
connection to the QA Portal
due to incorrect password.
- Enter the correct
password in the
analyzer's QA Portal
Communication Setup.
1196 ABL connected to
QA Portal
The analyzer is connected to
the QA Portal.
- No action required. For
information only.
1197 ABL disconnected
form QA Portal
The analyzer is disconnected
from the QA Portal.
- No action required. For
information only.
1198 ABL<>QA Portal
communication
error - XML
packet could not
be parsed
Communication error
between the analyzer and
the QA Portal.
- Contact IT engineer or
Radiometer service
representative.
1199 FTC program has
been retried
This message is found in the
Activity Log when a
measurement or calibration
activity has been retried due
to error.
- No action required. For
information only.
1200 Solution pack
empty
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1201 Solution pack
lifetime expired
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1202 Expiration date
reached
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1203 Lifetime in
analyzer
exceeded
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1204 No more
activities left
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals