10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-38
No. Message Interpretation Operator action
1223 Analyzer did not
connect at start-
up
The analyzer DMS has not
been able to establish
contact to the WS(M) at
start-up.
- Restart the analyzer.
- Contact Radiometer
service representative.
1224 Analyzer is
temporarily shut
down
Shown in the Activity Log
after temporary shutdown of
the analyzer.
- No action required.
1225 The sample is
older than a day
The time between sampler
draw time and aspiration is
larger than 1 day.
- Either sampler draw
time has been entered
incorrectly or time of
the analyzer is incorrect.
Change either to correct
the error.
1226 The sample age
is negative
The time between sampler
draw time and aspiration is
less than zero.
- Either sampler draw
time has been entered
incorrectly or time of
the analyzer is incorrect.
Change either to correct
the error.
1227 Correction for
bicarbonate
contains errors
from pH, pCO
2
Chloride is corrected for
bicarbonate, calculated from
pH and pCO
2
. Errors from
pH, pCO
2
results in this error
on chloride.
- No action required.
1228 Correction for
lactate contains
errors from K
+
,
Na
+
, Ca
2+
Lactate is corrected for ion
strength, calculated from K
+
,
Na
+
, Ca
2+
. Errors from K
+
,
Na
+
, Ca
2+
results in this error
on lactate.
- No action required.
1230 Inlet gasket
replaced
Shown in the activity log at
the time of a replacement.
- No action required.
1231 Inlet probe
replaced
Shown in the activity log at
the time of a replacement.
- No action required.
1232 Inlet connection
gasket replaced
Shown in the activity log at
the time of a replacement.
- No action required.
1233 Inlet cleaned Shown in the activity log at
the time when an inlet
cleaning was performed.
- No action required.
1234 Demonstration
software - not for
clinical purposes
1295 Activity has been
repeated due to
the following
reason:
T
his message is shown in the
activity log when an activity
is repeated automatically. It
lists the error and parameter
id that was the cause of the
repeat.
- No action required.
1296 Printer out of
paper
The printer is out of paper. A
new paper roll must be
inserted
- Insert a new paper roll

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals