No. Message Interpretation Operator action
1297 Printer is offline The printer is offline due to
either a bad or missing
power / USB connection
- Check the power
connection
- Check the USB
connection
- Contact Radiometer
service representative.
1298 Printer lid open The printer lid is open - Close the printer lid
1299 Rinse activity
repeated:
A rinse activity has been
repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1300 Calibration
activity
repeated:
A calibration activity has
been repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1301 QC activity
repeated:
A QC activity has been
repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1302 Startup activity
repeated:
A startup activity has been
repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1303 Activity
repeated:
An activity has been
repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1304 Calibration
activity repeated
A calibration activity has
been repeated. The following
entries in the log explain the
reason for the repeat.
- No action required.
1305 End of repeat
reason list
This message indicates the
end of repeat reasons. See
errors 1299-1304.
- No action required.
1306 Solution pack
manually
removed
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1307 Disk space less
than fifteen
percent
The disk space on the
analyzer is low.
- Delete some archives
to free up space on the
drive.
1308 Disk space less
than one percent
The disk space on the
analyzer is less than 1 %
- Free disk space. E.g.
deleting some archives
1309 Unable to start
FTC activity -
FTC activity in
progress

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals