10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-40
No. Message Interpretation Operator action
1310 Response error Sensor (Metabolite) does not
work properly
- Replace sensor
1311 The analyzer chip
data could not be
read or written
It's not possible to read or
write data to the analyzer
chip
- Contact Radiometer
service representative.
1312 Export data logs
failed
The export data log
operation has failed.
-
Make sure the selected
export path exists.
- Make sure enough
space is available.
1313 Export data logs
done
The export data log
operation has completed
successfully.
- No action required.
1314 Sensor
temperature
error during rinse
Sensor temperature error
(substrate) during rinse
- Check sensor status
and replace, if
necessary.
1315 Cal backlog error
(pH)
Cal backlog error (pH),
leaping signals on rinse
- Perform rinse
1316 Cal backlog error
(pCO
2
)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1317 Cal backlog error
(pO
2
)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1318 Cal backlog error
(K)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1319 Cal backlog error
(Na)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1320 Cal backlog error
(Ca)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1321 Cal backlog error
(Cl)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1322 Cal backlog error
(Glu)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1323 Cal backlog error
(Lac)
Backlog unstable, leaping
signals on rinse
- Perform rinse
1324 Inhomogeneous
rinse solution (LS
sensors)
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1325 Sensor
thermistor
recalibrated
Show in activity log when a
recalibration of the sensor
thermistor has been
performed
- Information only

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals