Remote Control
R&S
®
FSVA/FSV
655Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MINimum <Level>
This command sets the minimum lower level limit for one power class. The unit is dBm.
The limit always start at – 200 dBm, i.e. the first lower limit can not be set. If more than
one power class is in use, the lower limit must equal the upper limit of the previous
power class.
Suffix:
<n>
.
irrelevant
<k> irrelevant
<Class> 1...4
the power class to be evaluated
Parameters:
<Level> <numeric_value>
*RST: -200 for class1, otherwise +200
Example:
CALC:LIM:ESP:PCL2:MIN -40 dBm
Sets the minimum power value of the second power class to -40
dBm.
Manual operation: See "PMin/PMax" on page 355
CALCulate<n>:LIMit<k>:ESPectrum:RESTore
This command restores the predefined limit lines for the Spectrum Emission Mask
measurement. All modifications made to the predefined limit lines are lost and the fac-
tory-set values are restored.
Suffix:
<n>
.
1...4
window
<k> irrelevant
Example:
CALC:LIM:ESP:REST
Resets the limit lines for the Spectrum Emission Mask to the
default setting.
CALCulate<n>:LIMit<k>:ESPectrum:VALue <Power>
This command activates the manual limit line selection and specifies the expected
power as a value. Depending on the entered value, one of the predefined limit lines is
selected.
Suffix:
<n>
.
1...4
window
<k> irrelevant
Remote Control – Commands