Remote Control
R&S
®
FSVA/FSV
654Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:LIMit[:STATe] <State>
This command defines which limits are evaluated in the measurement.
Suffix:
<n>
.
irrelevant
<k> irrelevant
<Class> 1...4
the power class to be evaluated
Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation: See "Used Power Classes" on page 354
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MAXimum <Level>
This command sets the upper limit level for one power class. The unit is dBm. The limit
always ends at + 200 dBm, i.e. the upper limit of the last power class can not be set. If
more than one power class is in use, the upper limit must equal the lower limit of the
next power class.
Suffix:
<n>
.
irrelevant
<k> irrelevant
<Class> 1...4
the power class to be evaluated
Parameters:
<Level> <numeric value>
*RST: +200
Example:
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
Manual operation: See "PMin/PMax" on page 355
Remote Control – Commands