Remote Control
R&S
®
FSVA/FSV
653Operating Manual 1176.7510.02 ─ 09
Parameters:
<Mode> AUTO | MANUAL
AUTO
The limit line depends on the measured channel power.
MANUAL
One of the three specified limit lines is set. The selection is
made with the "CALCulate:LIMit:ESPectrum subsystem"
on page 652 command.
*RST: AUTO
Example:
CALC:LIM:ESP:MODE AUTO
Activates automatic selection of the limit line.
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>[:EXCLusive] <State>
This command sets the power classes used in the spectrum emission mask measure-
ment. It is only possible to use power classes for which limits are defined. Also, either
only one power class at a time or all power classes together can be selected.
Suffix:
<n>
.
irrelevant
<k> irrelevant
<Class> 1...4
the power class to be evaluated
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation: See "Used Power Classes" on page 354
See "Add/Remove" on page 355
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:COUNt <NoPowerClasses>
This command sets the number of power classes to be defined.
Suffix:
<n>
.
irrelevant
<k> irrelevant
<Class> irrelevant
Parameters:
<NoPowerClasses> 1 to 4
*RST: 1
Example:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Remote Control – Commands