Remote Control
R&S
®
FSVA/FSV
918Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Calibration Frequency RF" on page 193
DIAGnostic<n>:SERVice:INPut:PULSed:MCFRequency <Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(DIAGnostic<n>:SERVice:INPut[:SELect] on page 919).
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> <numeric value>
*RST: 7 GHz
Example:
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
Usage: SCPI confirmed
Manual operation: See "Calibration Frequency MW" on page 193
DIAGnostic<n>:SERVice:INPut:PULSed:WBFRequency <Frequency>
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration.
This command is only available if the bandwidth extension option R&S FSV-B160 is
installed.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic<n>:SERVice:INPut[:SELect] on page 919.
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> 2 MHz | 4 MHz | 8 MHz | 16 MHz
If you define a frequency that is not available, the
R&S FSVA/FSV uses the next available frequency.
*RST: 16 MHz
Example:
DIAG:SERV:INP:PULS:WBFR 8 MHz
Defines a calibration signal frequency of 8 MHz.
Usage: SCPI confirmed
Manual operation: See "Calibration Frequency WB" on page 193
DIAGnostic<n>:SERVice:INPut:PULSed:WB2Frequency <Frequency>
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration.
Remote Control – Commands