Remote Control
R&S
®
FSVA/FSV
919Operating Manual 1176.7510.02 ─ 09
This function is only available for R&S FSVA instruments. It is independant of the
bandwidth extension option R&S FSV-B160.
Before you can use the command, you have to define the use of a calibration signal
with DIAGnostic<n>:SERVice:INPut[:SELect] on page 919.
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> 16 MHz | 32 MHz
*RST: 32 MHz
Example:
DIAG:SERV:INP:SEL WB2
DIAG:SERV:INP:PULS:WB2F 16 MHz
Defines an internal wideband calibration signal frequency of 16
MHz.
Usage: SCPI confirmed
Manual operation: See "Calibration Frequency WB2" on page 194
DIAGnostic<n>:SERVice:INPut[:SELect] <Signal>
This command selects the source of the signal.
You can feed in a signal from the RF input or the calibration signal.
Suffix:
<n>
.
irrelevant
Parameters:
<Signal> RF
RF input
CALibration
Calibration
MCALibration
Microwave calibration (for R&S FSVA/FSV 13 and 30 only)
WBCal
Broadband signal for calibration of the IF filter.
WB2Cal
Broadband signal for calibration of the IF filter (independantly of
the B160 option).
This option is only available for R&S FSVA instruments.
*RST:
RF
Example:
DIAG:SERV:INP CAL
Selects the calibration signal as the input signal
Manual operation: See "RF" on page 193
See "Calibration Frequency WB" on page 193
See "Calibration Frequency WB2" on page 194