Instrument Function
R&S
®
SMBV100A
335Operating Manual 1407.6062.32 ─ 08
"On"
Corresponds to the default state of this parameter.
The signal calculation starts simultaneously with the external trigger
event but because of the instrument’s processing time the first samples
are cut off and no signal is outputted. After elapsing of the internal pro-
cessing time, the output signal is synchronous to the trigger event.
"Off"
The signal output begins after elapsing of the processing time and starts
with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequences
with signal duration comparable with the processing time of the instru-
ment.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:EXTernal:SYNChronize:OUTPut
on page 509
Trigger Delay
Sets the trigger signal delay in samples on external triggering or on internal triggering via
the second path.
Sets the trigger signal delay in samples on external triggering.
This enables the R&S Signal Generator to be synchronized with the device under test or
other external devices.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger[:EXTernal<1|2>]:DELay on page 514
Trigger Inhibit
Sets the duration for inhibiting a new trigger event subsequent to triggering. The input is
to be expressed in samples.
Baseband Signal - Baseband Block