EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 351

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
334Operating Manual 1407.6062.32 ─ 08
Execute Trigger
Executes trigger manually. A manual trigger can be executed only when an internal trig-
ger source and a trigger mode other than "Auto" have been selected.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:EXECute on page 509
Trigger Source
Selects trigger source. This setting is effective only when a trigger mode other than
"Auto" has been selected.
"Internal"
The trigger event is executed by "Execute Trigger".
"External
(TRIGGER 1 /
2)"
The trigger event is executed with the aid of the active edge of an exter-
nal trigger signal.
The trigger signal is supplied via the TRIGGER connector.
The polarity, the trigger threshold and the input impedance of the
TRIGGER input can be set in the "Global Trigger/Clock Settings" dialog.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:SOURce on page 513
Sync. Output to External Trigger
(enabled for "Trigger Source" External)
Enables/disables output of the signal synchronous to the external trigger event.
For R&S SMBV instruments:
For or two or more R&S SMBVs configured to work in a master-slave mode for synchro-
nous signal generation, configure this parameter depending on the provided system trig-
ger event and the properties of the output signal. See table 5-11 for an overview of the
required settings.
Table 5-11: Typical Applications
System Trigger Application "Sync. Output to External Trig-
ger"
Common External Trigger event for
the master and the slave instru-
ments
All instruments are synchronous to
the external trigger event
ON
All instruments are synchronous
among themselves but starting the
signal from first symbol is more
important than synchronicity with
external trigger event
OFF
Internal trigger signal of the master
R&S SMBV for the slave instru-
ments
All instruments are synchronous
among themselves
OFF
Baseband Signal - Baseband Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals