EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 350

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
333Operating Manual 1407.6062.32 ─ 08
"Single"
The signal is generated only when a trigger event occurs. Then the
signal is generated once to the length specified at "Signal Duration".
Every subsequent trigger event (internal with "Execute Trigger" or
external) causes a restart.
SCPI command:
[:SOURce<hw>]:BB:ARB[:TRIGger]:SEQuence on page 514
Signal Duration Unit
Defines the unit for the entry of the length of the signal sequence to be output in the
"Single" trigger mode.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:SLUNit on page 512
Signal Duration
Defines the length of the signal sequence to be output in the "Single" trigger mode. The
unit of the entry is defined under "Signal Duration Unit". It is then possible to output delib-
erately just part of the signal, an exact sequence of the signal, or a defined number of
repetitions of the signal.
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:SLENgth on page 512
Running/Stopped
Displays the status of signal generation for all trigger modes. This display appears only
when signal generation is enabled ("State" On).
"Running"
The waveform is output; a trigger was (internally or externally) initiated
in triggered mode.
If "Armed_Auto" and "Armed_Retrigger" have been selected, genera-
tion of signals can be stopped with the "Arm" button. A new trigger
(internally with "Execute Trigger" or externally) causes a restart.
"Stopped"
The Waveform output is stopped. The instrument waits for an internal
or external trigger event to restart. ARB modulation is disabled (State
Off).
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:RMODe on page 511
Arm
Stops waveform output. This button appears only with "Running" signal generation in the
"Armed_Auto" and "Armed_Retrigger" trigger modes.
Signal output can be restarted by a new trigger (internally with "Execute Trigger" or
externally).
SCPI command:
[:SOURce<hw>]:BB:ARB:TRIGger:ARM:EXECute on page 509
Baseband Signal - Baseband Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A