64 57 498 D3543
308 D3543.201.01.05.02 08.2016
10Overview of all function keys Sirona Dental Systems GmbH
10.4Integrated torque control Operating Instructions INTEGO
10.4
Integrated torque control
Integrated torque control
Activation of the integrated torque control function
Calibrating the burr drive
A calibration must be performed each time you change or lubricate the
contra-angle handpiece.
The contra-angle handpiece is automatically checked during calibration.
This includes a measurement of motor current at different speeds to
assess the properties of the system.
Direction of rotation
Switches counterclockwise rotation ON/OFF
Turning on/off the automatic motor stop of the ApexLocator
When the
Apex Stop
key is pressed, the and + keys are displayed. The
automatic motor stop can be switched off or set to four different levels. If
the distance is 0, the motor does not stop until it reaches the physiological
apex. Please note that the distance values are not a metric length
measurement!
The preset motor stop position is displayed as a black triangle to the right
of the distance display under the text "Stop".
Auto-reverse function
When the set torque value is reached, the bur drive automatically
switches to counterclockwise.
If your treatment center is equipped with the ApexLocator option, you can
determine that the bur drive automatically stops at the physiological apex.
If the Auto-reverse function is switched on, the next time the foot pedal is
activated following a motor stop, the motor is switched to
counterclockwise rotation. When the root canal file is withdrawn, the bur
drive automatically switches back to clockwise rotation.
Apex acoustic signals
Switching the apex signal tones on and off. If the signal tones are
switched on, a beep is issued when the apex or a set motor stop position
is reached. If the signal tones are switched on, a beep is issued when the
apex or a set motor stop position is reached. For more information on the
acoustic signals during apex measurement, see the section "Acoustic
signals".
Auto-reverse function
When the set torque value is reached, the bur drive automatically
switches to counterclockwise.
Signal tones
Switching the signal tones on and off. If the Auto-reverse function is
activated, three acoustic signals are issued when the motor switches to
counterclockwise rotation.
Memory key
Saves all integrated torque control settings
Cal
Apex
Stop
0.5
Apex
Auto
Rev
Auto
Rev