3Unit description Sirona Dental Systems GmbH
3.9Dentist element Operating Instructions INTEGO
Counterclockwise rotation/user profile
With the motor removed: switching counterclockwise rotation on/off, see
"Setting the rotation on the EasyPad" [ 104].
With instruments in place: changing the user profile, see "Selecting a user
profile" [ 61].
With motor removed and torque control switched on: Switch the
AutoReverse function on/off when maximum torque is reached by
pressing and holding the key (> 2 s), see "Switching AutoReverse ON/
OFF" [ 111]. Press briefly (< 2 s) to switch counterclockwise rotation on/
off.
Depending on the context, an ongoing process can be stopped and
switched to the standard operating context, e.g. ending the autopurge
process.
Endodontics/Purge
With the motor removed: Switch the integrated torque control on/off
"Switching integrated torque control ON/OFF" [ 106]
With the ultrasonic handpiece removed: switching the endo mode on/off
(limiting power), see "Setting the intensity on the EasyPad" [ 130].
With motor removed and torque control switched on: activating the
calibration process (> 2 s), see "Calibrating the burr drive" [ 107]
With instruments in place: selecting the purge, autopurge, sanitize, and
stand-alone water supply functions
Display mode/clean
Display mode/clean
With motor removed and torque control switched on: Switching between
torque and speed display, see “Setting the speed” [ 110] in the section
“Integrated speed control”.
With instruments in place: Deactivates the complete user interface of the
dentist element. The instruments of the dentist and assistant elements
can no longer be activated. The SIVISION monitor and the camera
system are switched off. Pressing it again (> 3 s) reactivates the user
interface. This is used to clean the surface and protect against
interference from an external HF surgical unit so that no unwanted
functions can be accidentally triggered, see “Disinfecting user interfaces”
[ 209]. Electromagnetic disturbances can often be reduced, if the
external HF surgical unit is operated with a neutral electrode. As long as
the suction removal system is required with the external HF surgical unit
during the treatment, the suction handpiece must be removed from the
holder before blocking the treatment center. The suction unit remains
switched on until the block is removed and the suction handpiece is put
down again.
Function key
Starts and stops timer, see "Timer function" [ 136].
A/B
Purge
Clean
Fn