EasyManua.ls Logo

Svantek SVAN 979 - A.2.2. Read-Out the RT60 Results; A.2.3. Read-Out the RPM Results

Svantek SVAN 979
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
185
SVAN 979 User Manual - Appendixes
T time of the measurement (ccc value in seconds);
P the PEAK value (ccc the value in dB);
Q the P_P value (ccc the value in dB);
M the MAX value (ccc the value in dB);
R the RMS value (ccc the value in dB);
H the VDV value (ccc the value in dB);
C the cycle of the measurement (ccc cycle number);
F date of the measurement (ccc date in the format yyyy-mm-dd where yyyy year, mm month, dd
day);
G time of the measurement (ccc time in the format hh:mm:ss where hh hour, mm minute, ss
second).
A.2.2. Read-out the RT60 results
In the case of RT60 function there are additional codes:
#2,EDT; this function allows you to read the current EDT results.
#2,T20; this function allows you to read the current RT20 results.
#2,T30; this function allows you to read the current RT30 results.
The function’s answer is in the one of the following format:
#2,<type>,0,<status>; when results are not available
#2,<type>,1,<results>; when results are available.
<type> accepts the following values: EDT/T20/T30
<status> is the status of device (0 no results, 1 waiting for trigger, 2 measurement in progress, 3
calculations)
<results> is the full list of frequencies with corresponding results in the format:
freq1:value1,freq2:value2,…,freqN:valueN
The exemplary answer:
#2,T30,1,50.0Hz:---,63.0Hz:---,80.0Hz:---,100Hz:0.48s,125Hz:---
,160Hz:0.31s,200Hz:0.61s,250Hz:0.62s,315Hz:0.65s,400Hz:0.56s,500Hz:0.66s,630Hz:0.81s,800Hz:0.69s
,1.00k:0.74s,1.25k:0.75s,1.60k:0.70s,2.00k:0.69s,2.50k:0.58s,3.15k:0.63s,4.00k:0.61s,5.00k:0.60s,6.30k:
0.45s,8.00k:0.35s,10.0k:0.26s,12.5k:0.20s,16.0k:0.25s,20.0k:0.18s,TOT.A:0.70s,TOT.C:0.72s,TOT.Z:0.8
2s;
A.2.3. Read-out the RPM results
In the case of activation RPM option there are additional codes:
#2,t,rps; The instrument sends the values of results in the format defined as follows:
#2,t,rps,<ROT>,<RMAX>,<RMIN>;
#2,t,rpm; The instrument sends the values of results in the format defined as follows:
#2,t,rpm,<ROT>,<RMAX>,<RMIN>;
t optional type of the results (a averaged, i instantaneous)

Table of Contents

Related product manuals