EasyManua.ls Logo

Svantek SVAN 979 - D.1.2 Definitions and Formulas of SLM Results

Svantek SVAN 979
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
357
SVAN 979 User Manual - Appendixes
D.1.2 Definitions and formulas of SLM results
The instrument calculates the sound measurement results for three profiles. The calculation flow diagram for one
profile is presented below:
Ovl
Percentage of the overloaded input signal, which occurred during the current time period of the
measurement (T)
Lpeak
Peak sound level is calculated for the given T
( )
=
2
0
2
W
p
tp
maxlog10Peak
T
Spl or LXY
Maximal value of the frequency and time-
weighted sound pressure level for the last
second of the measurement
( )
=
2
0
2
W
1
p
tp
maxlog10Spl
T
Lmax
Maximal value of the time-weighted sound
pressure level for current time period of the
measurement (T). The Max result for the 1
second period is equal to the value of the Spl
result
( )
=
2
0
2
W
p
tp
maxlog10Max
T
Lmin
Minimal value of the time-weighted sound
pressure level for current time period of the
measurement (T)
( )
=
2
0
2
W
p
tp
minlog10Min
T
Leq
(equivalent
sound level)
Time-averaged sound level for current time
period of the measurement (T)
( )
( )
=
T
0
2
0
dtptr
T
1
log10Leq
p
2
(t)
τ = Fast
τ = Slow
τ = Imp
p
w
(t)
p(t)
p
2
w
(t)
Lpeak
Lmax, Lmin,
Spl, LTeq
Leq,
SEL,
L(den),
LEPd

Table of Contents