EasyManua.ls Logo

Svantek SVAN 979 - Calibration by Sensitivity (Vibration Signal); Calibration by Measurement (Acoustic Signal)

Svantek SVAN 979
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
SVAN 979 User Manual
2. Set the Sensitivity of the microphone taken from its calibration certificate using the / key and then
press <ENTER>. The calibration factor (Cal. Factor) is calculated every time after pressing the / ► key,
in relation to the nominal value of 50.00 mV/Pa.
For microphones with sensitivity higher than 50.0 mV/Pa the calibration factor
will always be negative.
For microphones with sensitivity lower than 50.0 mV/Pa the calibration factor will
always be positive.
The lowest available value of the sensitivity that can be introduced is equal to
50.0 µV/Pa (it conforms to the calibration factor equal to 60.0 dB) and the highest
value is equal to 50.0 V/Pa (calibration factor is equal to -60.0 dB).
3. To confirm the new calibration factor, press <ENTER>, or to return to the Calibration list without changes,
press <ESC>.
3.3.3 Calibration by Sensitivity (Vibration signal)
The calibration by using the accelerometer’s published sensitivity information
can be performed in the following way:
1. Select the Calibr. by Sensitivity position in the Calibration list and press
<ENTER>.
2. Set the Sensitivity of the accelerometer taken from its calibration certificate
using the / ► key.
The calibration factor (Cal. Factor) is calculated every time after pressing the
/ ► key, in relation to the nominal value of 10.0 mV/ms
-2
.
For accelerometers with sensitivity higher than 10.0 mV/ms
-2
the calibration
factor will always be negative.
For accelerometers with sensitivity lower than 10.0 mV/ms
-2
the calibration factor
will always be positive.
The lowest available value of the sensitivity that can be introduced is equal to
10.0 µV/ms
-2
(it conforms to the calibration factor equal to 60.0 dB) and the
highest value is equal to 10.0 V/ms
-2
(calibration factor is equal to -60.0 dB).
3. To confirm the new calibration factor, press <ENTER>, or to return to the Calibration list without changes,
press <ESC>.
3.3.4 Calibration by Measurement (Acoustic signal)
Calibration by measurement of the acoustic signal can be done in the following
way:
1. Select the Calibr. by Measurement position in the Calibration list and press
<ENTER>.

Table of Contents

Related product manuals