After calibration measurement stop, the Calibration Result (the measured
reference signal without calibration factor correction) and the
New Calibration Factor (difference between the Calibration Level and the
Calibration Result, calculated in dB) are displayed.
3. Press <ENTER> to save the new calibration factor (Enter to Confirm),
press <ESC> to reject it. In both cases the instrument exits the Calibration
by Measurement screen.
4. Detach the calibrator from the microphone.