8-4 Sysmex XP-300
Display and Output of Analysis Results
Revised May 2012
Description of the analysis result screens
[Out/Del]: Prints the analysis results. For details, see “Output of analysis result
(latest sample)” in “8.1 Latest sample (analysis result screen)”.
[M. Discri.]: Relocates particle distribution discrimination in the histogram and re-
calculates the data. For details, see “Manual discrimination” in
“8.1 Latest sample (analysis result screen)”.
[ID Edit]: Edits the sample ID number. For details, see “Editing the sample ID” in
“8.1 Latest sample (analysis result screen)”.
[←] [→]: Switches the analysis result screens (first - fourth) in display order. For
details, see “8.1 Latest sample (analysis result screen)”.
Sample ID: Displays the sample ID number.
Analysis mode: Displays the analysis mode for the sample.
Displays [WB] for the whole blood mode, and [PD] for the pre-diluted
mode.
Date and time of analysis: Displays the date and time when the analysis result was obtained.
Operator ID: Displays the operator ID of the analysis.
Analysis data: Displays the analysis data of each parameter.
Meanings of signs displayed on the left of the analysis data are as
follows:
If an analysis error has occurred and a value is not available, one of the
following is displayed:
Sign Explanation
[!] Value is out of the linearity limit.
[+] Result exceeds the upper patient limit.
[-] Result exceeds the lower patient limit.
[∗] Result is unreliable.
The values for [+] and [-] patient limits can be set by the customer.
Check the “Patient Limits” in “11. Instrument Setup”.
Display Explanation
[+++.+] Value exceeds display range.
[∗∗∗.∗] Value could not be calculated because of analysis error.
At this time, the analysis error flag, [ERROR] (inverse in
coloration display) appears.
[---.-]
Value could not be calculated due to data error, or
volume distribution analysis parameter is not displayed
when the analysis was performed in pre-diluted mode.