3D OCT-1 PC SOFTWARE LICENSE TERMS
10. When the OCT instrument and "FastMap (PC)" have already shifted to the photography mode, the
system does not take a picture occasionally though the operator presses the photography trigger.
In this case, it is probable that LAN is disconnected temporarily. After checking the LAN connection,
finish "FastMap (PC)" to finish the procedure. Then, restart the procedure with [Capture Data].
11. After pressing the [Capture Data] button, the system does not respond for 30 - 40 seconds. Then,
"3D OCT is not ready" is displayed. The probable causes are shown below.
The OCT instrument is not turned on.
The OCT instrument is turned on but the application program is in the start process.
The application program has started but the LAN cable is not connected.
After checking the LAN connection, turn on the OCT instrument. Make sure that the application pro-
gram in the OCT instrument has started and then press the [Capture Data] button.
12. In the case of "Radial" and "Cross", the B-Scan image is sometimes displayed reversely on the analy-
sis report (New report) and "RawData" report (Old report).
The reason is that the B-Scan image direction is modified on the analysis report (New report) to make
diagnosis easier. There is no problem.