3D OCT-1 PC SOFTWARE LICENSE TERMS
5. When analyzing, please do not change the following setting.
"Volume Scale" and "B-Scan Scale" in "Display" of "Options"
6. It is necessary to install "Internet Explorer Version 8 (IE8)" or "Internet Explorer Version 9 (IE9)" in the
personal computer for 3D OCT. If any other browser except these is used, the report cannot be output
properly from time to time.
7. If you perform the following actions while the report is displayed, the report is not displayed correctly
from time to time.
• Turning on/off (connecting/disconnecting) the printer
• Changing the setting for the printer that is usually used on Windows
In this case, return to the "Viewer" screen and display the report again.
8. Sometimes the "FastMap (PC)" grabber timer does not count though the counter on the OCT instru-
ment operates in the fundus grabber mode.
In this case, it is probable that LAN is disconnected temporarily. After checking the LAN connection,
finish "FastMap (PC)" to finish the procedure. Then, restart the procedure with [Capture Data].
9. Sometimes the "FastMap (PC)" grabber is not activated though the OCT instrument takes a picture in
the fundus mode.
In this case, it is probable that LAN is disconnected temporarily. After checking the LAN connection,
finish "FastMap (PC)" to finish the procedure. Then, restart the procedure with [Capture Data].