EasyManuals Logo

YOKOGAWA ADMAG TI Series User Manual

YOKOGAWA ADMAG TI Series
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #173 background imageLoading...
Page #173 background image
<6. Parameter Lists>
172
IM 01E21A02-03EN
Relative
Index
Index Parameter Name Default Value
Write
Mode
Description
41 1041 ITK_VER 6 -
VersionnumberoftheFFcertication
test(interoperabilitytest)byFieldbus
Foundation applied to this product.
42 1042
COMPATIBILITY_
REV
2 -
Indicates the lower revision compatible
with the device DevRev.
43 1043 CAPABILITY_LEV 0 -
Indicates capability level inside the
device.
44 1044 FD_VER 1 -
Indicates the major version value of
diagnosisspecicationofthedevice.
45 1045 FD_FAIL_ACTIVE 0x00000000 -
Parameterassignedto“Failed”bythe
NAMURNE-107classication.For
detailsaboutbitassignment,referto
Field Diagnostic.
46 1046
FD_OFFSPEC_
ACTIVE
0x00000000 -
Parameterassignedto“O󰀨
Specication”bytheNAMURNE-107
classication.Fordetailsaboutbit
assignment,referto■FieldDiagnostic.
47 1047
FD_MAINT_
ACTIVE
0x00000000 -
Parameterassignedto“Maintenance”
bytheNAMURNE-107classication.
For details about bit assignment, refer to
■FieldDiagnostic.
48 1048
FD_CHECK_
ACTIVE
0x00000000 -
Parameterassignedto“CheckFunction”
bytheNAMURNE-107classication.
For details about bit assignment, refer to
■FieldDiagnostic.
49 1049 FD_FAIL_MAP 0xFC000000 AUTO
SpeciesabittoassigntoFD_FAIL_
ACTIVE,parameterindicating“Failed”,
among 32-bit alarms appearing on
FD_SIMULATE.DiagnosticValue. For
detailsaboutbitassignment,referto
Field Diagnostic.
50 1050
FD_OFFSPEC_
MAP
0x00003800 AUTO
SpeciesabittoassigntoFD_
OFFSPEC_ACTIVE, parameter
indicating“O󰀨Specication”among32-
bit alarms appearing on FD_SIMULATE.
DiagnosticValue. For details about bit
assignment,referto■FieldDiagnostic.
51 1051 FD_MAINT_MAP 0x000003E0 AUTO
SpeciesabittoassigntoFD_
MAINT_ACTIVE, parameter indicating
“Maintenance”,among32-bitalarms
appearing on FD_SIMULATE.
DiagnosticValue. For details about bit
assignment,referto■FieldDiagnostic.
52 1052 FD_CHECK_MAP 0x01FF8008 AUTO
SpeciesabittoassigntoFD_
CHECK_ACTIVE,parameterindicating
“CheckFunction”,among32-bit
alarms appearing on FD_SIMULATE.
DiagnosticValue. For details about bit
assignment,referto■FieldDiagnostic.
53 1053 FD_FAIL_MASK 0x00000000 AUTO
Speciesabitnottonotifytothehost
among32-bit“Failed”alarmsappearing
on FD_FAIL_ACTIVE. For details
aboutbitassignment,referto■Field
Diagnostic.
54 1054
FD_OFFSPEC_
MASK
0x00000000 AUTO
Speciesabitnottonotifytothehost
among32-bit“O󰀨Specication”alarms
appearing on FD_OFFSPEC_ACTIVE.
For details about bit assignment, refer to
■FieldDiagnostic.
55 1055
FD_MAINT_
MASK
0x00000000 AUTO
Speciesabitnottonotifytothehost
among32-bit“Maintenance”alarms
appearing on FD_MAINT_ACTIVE. For
detailsaboutbitassignment,referto
Field Diagnostic.

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals