<5. Functions>
97
IM 01E21A02-03EN
NE
107
Error Message
Error Description Countermeasure FD bit
Display
F
OUNDATION
Fieldbus
C 101:Param restore run
Parameter restore
running
The restore function
of the parameter was
executed.
- 4 15
N 102:Disp over
Display over
warning
The number of digits
available for display
exceeded the limit.
Check and change the
display parameter.
4 14
N 103:SD size warn
microSD card size
warning
Free space of
the microSD card
decreased to less than
10%.
microSD card may run
out of memory space.
4 13
M 104:Bkup incmplt
Parameter backup
incomplete
Parameter backup
failed.
Retry parameter
backup.
4 12
S 105:SD mismatch
microSD card
mismatch
Mismatch of the
microSD card was
detected.
Change the microSD
card.
4 11
M 106:SD removal ERR
microSD card
removal procedure
error
Removal of the
microSD card failed.
Remove the microSD
card in appropriate
procedures.
4 10
N 131:Trans mismatch
Transmitter type
mismatch
Mismatch of sensor
and transmitter was
detected.
Contact Yokogawa
service center.
3 0
 Information
The device works normally and measurement is also normal. These messages are just reference
information.
NE
107
Error Message
Error Description Countermeasure FD bit
Display
F
OUNDATION
Fieldbus
N 120:Watchdog Watchdog
Error of Watchdog
timer was detected.
Contact Yokogawa
service center.
4 7
N 121:Powero󰀨 Powero󰀨
Power-o󰀨was
detected.
- 4 6
N 122:Inst power FAIL Instant power failure
Instantaneous power
failure was detected.
- 4 5
N 123:Param bkup run
Parameter backup
running
Parameter backup
function is running.
- 4 4
N 124:Data log run
Data logging
running
Data-logging function is
running.
- 4 3
N 133:G/A mismatch G/A mismatch error
As a result of self-
test, inconsistency of
internal circuit was
detected.
Contact Yokogawa
service center.
5 21
N - Write Unlocked
Write lock function
inoperative(Parameter
canbewritten).
Make Write Lock of
resource block Locked.
1 30
N -
Hard Write Lock SW
OFF
Hardware write
protection switch is
turnedo󰀨(writing
enabled).
Turn on the hardware
write protection switch.
1 29
N - Write Locked
Write lock function
operative(Parameter
cannotbewritten).
Make Write Lock
of resource block
Unlocked.
1 28
N -
Hard Write Lock SW
ON
Hardware write
protection switch is
turnedon(writing
disabled).
Turno󰀨thehardware
write protection switch.
1 27
N 248:Simulate SW ON
Simulation Switch
ON
Simulation switch is
turned on.
Turno󰀨thesimulation
switch.
1 23

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals