3-26
IM MW100-01E
Scale Input Methods
After selecting the number of digits after the decimal point for the upper or lower limit
value from the list, enter
the number you wish to set in the input box. If the number of digits after the decimal
place set in the input box is
larger than the specified number of digits, it will be rounded. (For example, if the decimal
place is set to 2 and
you enter a value of 95.006, it will be rounded to 95.00.)
Desired Scale Decimal Point Position Number Input
0.00 to 100.00 2 Lower limit: 0
Upper limit: 10000
10.0 to 500.0 1 Lower limit: 100
Upper limit: 5000
–6.000 to 4.500 3 Lower limit: –6000
Upper limit: 4500
Example of a filled in screen.
Filter and Thermocouple Settings
You can set filters for measurement channels. Also, when the input type is set to
thermocouple (TC), you can set the burnout/RJC behavior. This setting is allowed even if
the input type of the measurement channel is something other than thermocouple (TC),
but has no effect in that case.
From the top screen, click Channel Setting > Filter, Burnout, RJC Setting.
Setting the Filter Coefficient
1.
Select a coefficient in the Filter list.
2.
Click the Apply button. The setting changes take effect.
For information about filter coefficients, see section 2.9, “Measures against Noise on the
MW100 Data Acquisition Unit.”
3.5 Setting Measurement Conditions indstop (indstart Measurement Channel Settings)