Setting and Data acquisiton
3
Setting the Burnout
1.
Select the direction in which the measured value is set off the range when a
burnout detection occurs from the Burnout list.
2.
Click the Apply button. The setting changes take effect.
For a description of the burnout detection behavior, see “Burnout” in section 1.3.
Reference Junction Compensation
1.
Select the type of reference junction compensation in the Type box under RJC.
2.
Enter a reference junction compensation voltage in the Voltage[uV] box. Set
when reference junction compensation is set to External.
3.
Click the Apply button. The setting changes take effect.
For the setting range of the RJC, see “RJC” in section 5.2.
Setting Up and Executing Strain Input Initial Balancing
If the measurement channel is strain input, you can enter settings for executing initial
balancing. This setting is allowed even if the measurement channel is something other
than strain input, but has no effect in that case.
1.
From the top screen, click Channel Settings > Strain Input Setting.
2.
Select the channel group you wish to set from the Channel List list.
3.
Turn On the channel on which you wish to perform initial balancing in the Initial
Balancing list.
4.
Click the Apply button to finalize.
5.
Execute the Initial balance items in the Status information of the system settings.
For executing the initial balancing, see “Processing and Operation” in “Status Information
and Processing” in section 3.3.
For a description of initial balancing, see “Initial Balancing (Unbalance Adjustment)” in
section 1.7.
3.5 Setting Measurement Conditions indstop (indstart Measurement Channel Settings)