1-37
IM MW100-01E
Explanation of Functions
1
1.7 Functions of the 4-CH, Medium-Speed Strain
Input Module
This module allows up to four inputs of measurement from strain gauges and strain
gauge type sensors at a minimum measurement interval of 100 ms.
-B12, -B35 -NDI
Input terminal
(clamp terminal)
Terminal cover
NDIS terminal
Measurement Input Types
Measurement Input Type Notation
No measurement SKIP
Strain STR
Measurement Range
Strain Input
Measurement Range Type Notation Rated Measurement Range
2000 µSTR 2000 uSTR –2000.0 to 2000.0 µSTR
20000 µSTR 20000 uSTR –20000 to 20000 µSTR
200000 µSTR 200000 uSTR –200000 to 200000 µSTR
Measurement Interval, Integration Time, and Filter
You can select from the following measurement intervals for this module.
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 60 s
The integral time and types of filters applied vary depending on the measurement
interval. For details, see section 2.9, “Measures against Noise on the MW100 Data
Acquisition Unit.”
Measurement Synchronization
Since measurement is sequential, measurement on each channel is not synchronized.