EasyManuals Logo

YOKOGAWA MW100 User Manual

YOKOGAWA MW100
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
1-26
IM MW100-01E
RS-422A/485 Communication (/C3 Option)
In a multi-drop, four-wire configuration, up to thirty-two units can be connected. A
dedicated protocol and the ModbusRTU protocol are supported.
Using communication commands, you can send and receive settings, and measured and
computed values.
Modbus Master Function
The MW100 can connect to a Modbus slave device and load measured data. The
operation is the same as the Modbus client function. Target devices (Modbus slaves)
are identified with addresses.
MW100(master)
MW100
Data from slave device
Data to slave device
Modbus slave device
Serial communication
Modbus Slave Function
Modbus master devices can connect to other MW100s acting as Modbus slaves and
read from or write to their internal registers. The MW100 registers are the same as
those for the Modbus server function.
MW100(slave)
MW100
Data to master device
Data from master
device
Modbus master device
Modbus slave device
Serial communication
For instructions on connections, see section 2.7, “Connections Using the RS-422A/485
Interface (/C3 Option).”
RS-232 Communication (/C2 Option)
This is a point-to-point system. A dedicated protocol and the Modbus/RTU protocol are
supported.
Using communication commands, you can send and receive settings, and measured
and computed values.
The MW100 operates as the Modbus master or slave. For information about the
operation, see “RS-422A/485 Communications (/C3 Option).”
For instructions on connections, see section 2.8, “Interface Connections (/C2 Option).”
1.3 Functions of the Main Module

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MW100 and is the answer not in the manual?

YOKOGAWA MW100 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMW100
CategoryData Loggers
LanguageEnglish

Related product manuals