Running a Single Test Running a Single Test
10-2 PN: 13000-00205 Rev. N MT8852B Operation Manual
7. Use the key to navigate to any test conditions that require changing. Press the
key to toggle the settings or to enable entry of a new value.
8. Press the [Limits] soft key to display the test limits for the selected test. The limits are
set by default to match the requirements set out in the Bluetooth specification.
9. Press the [Save] soft key to save the test conditions.
10. Press to run the script once, or press to run the test script in
accordance with the instrument-wide script settings outlined in chapter 5.
11. If required, press the [Abort] soft key to terminate the script immediately. Press
again to stop a continuous script.
12. The test results display.
Note
Refer to Appendix B for an alphabetic listing of all test conditions and their
meanings.
Changes made to the test conditions are not automatically carried over to the
settings in the selected script. When the user returns to the Script or Config pages,
a message displays asking whether the changes should be saved to the script.
Note
Changes made to the limits are not automatically carried over to the settings in the
selected script. When the user returns to the Script or Config pages, a message
displays asking whether the changes should be saved into the script.