Running a Script in Standard Mode Low Energy Tests
MT8852B Operation Manual PN: 13000-00205 Rev. N 7-51
Sensitivity Test
After sending a test control to prepare the EUT, the MT8852B sends BLE reference packets
to the EUT. The number of packets received at the EUT is counted and this data is then read
by the MT8852B over the HCI or 2-Wire connection.
The table below details the test conditions and the associated default settings. Permitted
variations from the default setting are detailed in the “User Range” column.
Results
1. Position the cursor at "Sensitivity" in the Script menu.
2. Press the [Results] soft key.
The sensitivity results pages displays.
Figure 7-70. Low Energy Sensitivity Test
Table 7-21. Sensitivity Test Conditions
Parameter Default value User Range
Number of packets 1500 (440000 bits) 1 to 10,000
Test Tx power -70.0 dBm -90.0 dBm to 0.0 dBm
Dirty transmitter ON OFF, ON
Packet type BLE BLE, 2LE, BLR2, BLR8
Figure 7-71. Low Energy Sensitivity Results 1