Test Scripts Configuring a Script
MT8852B Operation Manual PN: 13000-00205 Rev. N 5-11
Configuring Instrument-Wide Script Settings
Configure the parameters that are applied to all scripts. These settings include items that
determine if and how tests are looped.
1. Press .
2. Press the key twice to select "All Scripts Setup".
3. Press .
The Scripts setup page displays.
4. Use the key to select the appropriate setting at “Loop stop on test fail”. If set to
“ON” the test is run repeatedly until a failure is recorded or the test is stopped.
5. Move the cursor to “Loop continuously” and use the key to select the appropriate
setting. If set to “ON”, the test is run continuously until stopped.
6. Move the cursor to “Loop count”, press the key, and use the keypad to enter the
number of times that the test will be looped. Press the [Enter] soft key after entering
the number.
7. Move the cursor to “Display frequency as" and use the key to select the
appropriate setting. This setting determines whether a channel displays in number
format (0 to 78) or in frequency format (MHz).
Figure 5-12. All Scripts Setup