Testing in Null Packet Mode Running a Script in Null Packet Mode
9-8 PN: 13000-00205 Rev. N MT8852B Operation Manual
Carrier Drift Test
The carrier drift test performs a frequency drift measurement between two defined sections of
the Null packet transmitted by the EUT. The average frequency of the preamble is measured
as described for the initial carrier frequency test. This is then compared to the average
frequency of the four trailer bits in the Null packet.
The table below details the test conditions and the associated default settings. Permitted
variations from the default setting are detailed in the “User Range” column.
Results
Standard and single packet modes
1. Position the cursor at "Carrier drift" in the Script menu.
2. Press the [Results] soft key.
The Carrier drift results page displays.
Figure 9-10. Carrier Drift Test
Table 9-5. Carrier Drift Test Conditions
Parameter Default value User Range
Hopping test mode Defined Defined, All, Any
Number of packets 10 1 to 10,000
Figure 9-11. Carrier Drift Results