Running a Script in Standard Mode Basic Rate Tests
MT8852B Operation Manual PN: 13000-00205 Rev. N 7-19
The Multi sensitivity results page displays.
Current BER: Displays the BER as measured in the last 10 returned packets.
Overall BER: Displays the BER for all the returned packets.
Current FER: Displays the FER as measured in the last 10 returned packets.
Overall FER: Displays the FER for all the returned packets.
3. Press the [1 of 3] soft key to move to the second page.
4. Press the [2 of 3] soft key to move to the third page.
FER due to CRC: Displays the number of packets returned that had a Cyclic
Redundancy Check (CRC) error. These packets are used in BER
calculations.
Figure 7-25. Multi Sensitivity Results 1
Figure 7-26. Multi Sensitivity Results 2
Figure 7-27. Multi Sensitivity Results 3