Running a Script in Standard Mode Basic Rate Tests
MT8852B Operation Manual PN: 13000-00205 Rev. N 7-25
Current BER: Displays the BER as measured in the last 10 returned packets.
Overall BER: Displays the BER for all the returned packets.
Current FER: Displays the FER as measured in the last 10 returned packets.
3. Press the [1 of 3] soft key to move to the second page.
4. Press the [2 of 3] soft key to move to the third page.
FER due to CRC: Displays the number of packets returned that had a Cyclic
Redundancy Check (CRC) error. These packets are used in
BER calculations.
FER due to length: Displays the number of packets returned with incorrect
payload lengths. This will also result in a CRC error. These
packets are used in BER calculations.
FER due to lost packets:A lost packet will result from the MT8852B under the
following conditions:
Not recognizing the sync word header as being correct.
The packet failing the Header Error Correction (HEC)
No packet received during the expected time slot.
NACK packet is received.
Figure 7-35. Max Input Power Results 2
Figure 7-36. Max Input Power Results 3