EasyManua.ls Logo

Baumuller b maXX BM5800 - Description of the Parameters

Baumuller b maXX BM5800
816 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Description of the Software Modules and Parameters
Parameter manual b maXX BM5800
Document No.: 5.16029.03
183
of 814
3
3.4.7.3 Description of the Parameters
123.1 Mode
The contents of the identification can be set with this parameter.
To bit 3: Mode of the adaption of the dead time compensation after the PWM frequency:
If the adaption is requested, the bit must be set before the measurement and then
must remain the same.
If the bit is changed, the dead time voltage measurement must be repeated.
If the adaption is activated, the dead time measurement values are converted to val-
ues corresponding to a PWM frequency of 8 kHz and subsequently stored in the
dead time correction table
Z123.15. The effective dead time voltage for the dead
123.7 Inductance measured FLOAT 0.0 10000 6.4 mH 1:1 X
123.8 Result current controller
Kp
FLOAT 0.0 1000 0.0 V/A 1:1 X
123.9 Result current controller
Tn
FLOAT 0.0 1000 0.0 ms 1:1 X
123.10 Parameters takeover DINT 0 15 0 1:1 X
123.12 Time constant of step
response
FLOAT 0 10000 0 ms 1:1 X
123.14 Normalized Error FLOAT 0 100 0 % 1:1 X
123.15 Dead time compensa-
tion table
FLOAT -100 100 0 V 1:1 X
123.35 Result Rr FLOAT 0 100 0 Ohm 1:1 X
Bit Meaning
0 Mode motor resistance and inductance measurement
0: No measurement
1: Measurement
1 Mode dead time measurement
0: No measurement
1: Measurement
2 Reserved
3 Mode of the adaption of the dead time compensation after the PWM fre-
quency:
0: Not activated
1: Activated
4 Motor diagnosis:
0: Not activated
1: Activated
The motor diagnosis will be initiated if only bit 4 is activated and all other
bits were deactivated.

Table of Contents

Related product manuals