System Error Codes | 285
Code Message What to do
217 Strip washer syringe motor
didnt find home opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
218, 228 Aspiration head motor didn’t
find home opto sensor
transitionAspiration head motor
didnt find autocal jig opto
sensor transition
Run self test. If error reoccurs, contact
BioTek TAC.
219, 229 RAD Y-axis motor didnt find
home opto sensor transitionRAD
Y-axis motor didnt find autocal
jig opto sensor transition
Run self test. If error reoccurs, contact
BioTek TAC.
220 Carrier X motor didnt find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
221 Carrier Y motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
300 Carrier X motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
301 Carrier Y motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
302 Dispense head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
303 Secondary Peri-pump motor
interlock safety switch open
Service Only.
Contact BioTek TAC.
304 Syringe A motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
305 Syringe B motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
306 Peri-pump Pump Cover is open Close the pump cover door and rerun the
protocol.
307 Strip washer motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
308 Aspirate head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.