286 | Appendix A: Error Codes
Code Message What to do
309 RAD Y-axis motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
400 Carrier X motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
401 Carrier Y motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
402 Dispense head motor failed
positional verify
Run self test. If error reoccurs, contact
BioTek TAC.
403 Secondary Peri-pump motor
failed positional verify
Service Only.
Contact BioTek TAC.
404 Syringe A motor failed
positional verify
Service Only.
Contact BioTek TAC.
405 Syringe B motor failed
positional verify
Service Only.
Contact BioTek TAC.
406 Peri-pump motor failed
positional verify
Service Only.
Contact BioTek TAC.
407 Strip washer syringe motor
failed positional verify
Service Only.
Contact BioTek TAC.
408 Aspirate head motor failed
positional verify
Service Only.
Contact BioTek TAC.
409 RAD Y-axis motor failed
positional verify
Service Only.
Contact BioTek TAC.
600-609 Specified motor currently in use Service Only.
Contact BioTek TAC.
700 -
70F
Invalid motor number specified Service Only.
Contact BioTek TAC.
900 Calibration failed. The
measured or calculated autocal
value is out of tolerance
Service Only.
Contact BioTek TAC.
A00 Invalid plate type selected The currently selected plate type is not
supported with the currently installed or
requested hardware. If this is not the case,
contact BioTek TAC.