MultiFlo FX Software Error Codes | 299
Error
Code
Description Help
6088 Invalid Shake/Soak options selected Service Only.
Contact BioTek TAC.
These codes indicate an unexpected
software error that cannot be fixed
without BioTek support.
6089 Invalid Shake Intensity selected
6090 Invalid Washer buffer selected
6091 Invalid Washer Aspirate Delay value
6092 Invalid Washer Aspirate Travel Rate
value
6093 Invalid Wash Cycles value
6094 Invalid Wash format selected
6095 Invalid Wash Sectors selected
6096 Wash Aspirate Delay value is
required.
6097 Syringe Dispense Volume must be an
integer.
6098 Peri-pump cannot run with the pump
cover open.
Close the pump cover door and rerun the
protocol.
6099 Peri-pump assembly not installed. Physically install the Peri-pump and/or
make sure the instrument settings reflect
the current state.
6160 Strip washer required to perform this
step, but not installed.
Make sure the protocol you are trying to
run is supported by your instrument's
devices.
6161
6162
Strip washer manifold and plate type
are incompatible.
Strip washer doesn't support this
plate type.
Make sure the specified plate type
matches the installed manifold and the
instrument's manifold setting correctly
identifies the installed manifold.
6165 This Peri-pump does not support
single well dispensing.
An external Peri-pump is required for
RADtechnology cassettes. Edit the
protocol to use the correct Peri-pump,
i.e. protocols designed for a Primary
pump cannot be run on a Secondary
pump.
6166 This instrument does not support RAD
technology dispensing.
Make sure the protocol you are trying to
run is supported by your instrument's
devices.