ICS-3000 Ion Chromatography System
104 Doc. 065031-04 1/08
2.25.1 Predictive Performance
The Predictive Performance feature of the TC provides various functions
for estimating the lifetime of replaceable parts and for planning and
recording service and qualification procedures.
A partial list of the TC Predictive Performance functions is provided
below. Refer to the Chromeleon or Chromeleon Xpress Help for details.
• Monitoring of the total cooling and heating workload performed by
the compartment and the total operating hours since initial operation
• Monitoring of the injection valve usage
• Reminders for service and qualification periods, including display of
warning messages when limits are approaching or overdue
Predictive Performance commands and parameters are available in the
Commands dialog box.
Do not touch any metal parts inside the TC while it is heating up or
after it reaches the set point temperature. Wait for the compartment to
cool down before servicing any parts.
Ne touchez à aucune des pièces métalliques à l'intérieur du TC
pendant qu'il chauffe ou après qu'il ait atteint la température de
consigne. Attendez que le compartiment soit refroidi avant de réparer
les pièces.
Berühren Sie im TC keine metallischen Teile, während dieses
aufgeheizt wird oder nachdem die Solltemperatur erreicht ist. Warten
Sie, bis die Kammer abgekühlt ist, ehe Sie Wartungsarbeiten
durchführen.