Doc. 065031-04 1/08 165
DC Routine Maintenance
This section describes routine maintenance procedures for the ICS-3000
Detector/Chromatography Module (DC) that users may perform. All other
maintenance procedures must be performed by Dionex personnel.
7.9 DC Daily Maintenance
Check the DC components for leaks or spills. Wipe up spills. Isolate and
repair leaks (see Section 8.26
). Rinse off any dried eluent with ASTM filtered,
Type I (18 megohm-cm) deionized water.
Check the waste container and empty as needed.
NOTE If you are using an electrochemical detector, also see the
special maintenance considerations that are described in
Section 4.1.2
.
7.10 DC Weekly Maintenance
Check liquid lines for crimping or discoloration. Relocate any pinched lines.
Replace damaged lines.
Check for excess condensation inside the DC. Condensation may build up
when the DC is operated in a highly humid environment and the majority of
the applications are run at cold temperatures. To remove condensation, run the
DC at its maximum temperature for at least 48 hours for a dual-zone DC, or
72 hours for a single-zone (shared) DC.
7.11 DC Periodic Maintenance
Calibrate the CD cell (every 6 months).
Inspect the door seal for signs of damage. A defective seal impairs the
performance of the instrument. If the seal is defective, contact Dionex
Technical Support.