ICS-3000 Ion Chromatography System
282 Doc. 065031-04 1/08
9.11 Isolating a Restriction in the Liquid Lines
A restriction in the liquid plumbing will cause excessive system backpressure.
1. Begin pumping eluent through the system (including the columns) at the flow
rate normally used.
2. Work backward through the system, beginning at the flow cell exit. One at a
time, loosen each fitting and observe the pressure. The connection at which
the pressure drops abnormally indicates the point of restriction. If the EluGen
cartridge is the source of the high backpressure, replace the outlet frit as
instructed in the EluGen cartridge manual. The manual is included on the
Dionex Reference Library CD-ROM (P/N 053891).
3. If the restriction has caused such high pressure that the system cannot be
operated, you must work forward through the system, adding parts one at a
time, until an abnormal pressure increase (and hence, the restriction) is found.
9.12 Replacing the EluGen Cartridge
Replace the EluGen cartridge when the cartridge is expended or when it
leaks.
The EluGen cartridge replacement procedure consists of:
Removing the old EluGen cartridge (see Section 9.12.2)
Disposing of the old EluGen cartridge (see Section 9.12.3)
Installing the new EluGen cartridge (see Section 9.12.4)
Conditioning the new EluGen cartridge (see Section 9.12.5)
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), a corrosive acid (MSA), or a concentrated
K
2
CO
3
solution. Wear protective eyewear and gloves when handling
the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), un acide corrosif (MSA), ou une solution
concentrée de K
2
CO
3
. Porter des lunettes et des gants protectives en
manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), eine korrodierende Säure (MSA) oder eine konzentrierte
K
2
CO
3
-Lösung. Tragen Sie daher beim Umgang mit Kassette eine
Schutzbrille und Handschuhe.

Table of Contents

Related product manuals