Doc. 065031-04 1/08 167
TC Routine Maintenance
This section describes routine maintenance procedures for the ICS-3000 Thermal
Compartment (TC) that users may perform. All other maintenance procedures
must be performed by Dionex personnel.
7.13 TC Daily Maintenance
• Check the TC components for leaks or spills. Wipe up spills. Isolate and
repair leaks (see Section 8.31
). Rinse off any dried eluent with ASTM filtered,
Type I (18 megohm-cm) deionized water.
• Check the liquid level in the waste container and empty as needed.
7.14 TC Weekly Maintenance
• Check liquid lines for crimping or discoloration. Relocate any pinched lines.
Replace damaged lines.
Neutralize acidic and caustic wastes before disposal. Dispose of all
wastes in accordance with local regulations.
Neutralisez les déchets acides ou caustiques avant de les jeter. Jetez
les déchets aux règlements locaux.
Neutralisieren Sie säurehaltige und ätzende Abfälle vor ihrer
Entsorgung. Entsorgen Sie alle Abfälle entsprechend den lokalen
Bestimmungen.