EasyManua.ls Logo

Dionex ICS-3000 - DC Specifications; Electrical; Environmental; Physical

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. 065031-04 1/08 341
DC Specifications
A.11 Electrical
A.12 Environmental
A.13 Physical
Main Power
Requirements
90 to 265 VAC, 47 to 63 Hz (Auto-sensing power supply; no
manual voltage or frequency adjustment required)
Typical input power: 190 W
Maximum line draw: 9.2 A at 110 VAC
Fuse
Requirements
Two IEC 60127-2 slow-blow fuses (P/N 954772) rated at 10 A,
250 V
Operating
Temperature
(Ambient)
4 to 40 °C (40 to 104 °F)
Humidity
(Ambient)
5 to 95% relative humidity, noncondensing
Dimensions Height: 40.8 cm (16 in)
Width: 44 cm (17.5 in)
Depth: 54.9 cm (21.6 in)
Weight 38 kg (84 lb)