EasyManua.ls Logo

Dionex ICS-3000 - High Background

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 • Troubleshooting
Doc. 065031-04 1/08 187
with pressurized air or nitrogen. If the electrode is pitted, polish the
electrode (see Section 9.24.3
).
3. Check the value of the reference potential shift. See
Using a Digital
Voltmeter to Determine Reference Potential Shift” on page 226. If the
value fluctuates by more than 30 mV, electrode passivation may occur
because potentials that are too high are being applied. Recalibrate the
reference electrode (see Section 9.24.5
). If the problem persists, replace
the reference electrode (see Section 9.24.4
).
Analog Output Option
Analog output range set too high
Select a more sensitive analog output range.
8.11 High Background
CR-TC is contaminated
Clean the trap column as instructed in the CR-TC manual. The manual is
included on the Dionex Reference Library CD-ROM (P/N 053891).
Wrong eluent
Check that you are using the correct eluent for your application. For ED
detectors, verify that the pH readout is correct for your eluent.
Background signal not offset from detector signal
Before injecting sample, allow the background signal to equilibrate, and then
press Autozero on the Control panel in Chromeleon or Chromeleon Xpress.
CD Only
Background not suppressed by suppressor
Verify that the suppressor is turned on and the current is set to the correct
value. Refer to the suppressor manual for additional troubleshooting
guidance. Suppressor manuals are included on the Dionex Reference Library
CD-ROM (P/N 053891).
Regenerant fails to suppress background
Use a higher regenerant flow rate. Refer to the suppressor manual for the
suggested regenerant flow rate for your application.

Table of Contents

Related product manuals