• (Integrated Amperometry mode) Excessive number or length of
integration intervals and/or incorrect potential for the integration
Verify that the length and potential of the integration interval is correct (refer
to the column manual for the settings required for your application). Column
manuals are included on the Dionex Reference Library CD-ROM
(P/N 053891).
• (DC Amperometry and Integrated Amperometry modes) Amperometry
cell working electrode shorted to counter electrode
Clean the working electrode with water and dry with pressurized air or
nitrogen.
Remove any precipitate on the counter electrode by cleaning the spot directly
opposite the working electrode with a lint-free tissue.
• (DC Amperometry and Integrated Amperometry modes)—Leak between
gasket and electrode, or between gasket and cell body
Remove any precipitate with water and a lint-free tissue and then install a new
gasket (see Section 9.24.2
).
Turn the PEEK yoke knob (see Figure 9-33
) a full 360 degrees to completely
engage the knob.