9.24.3 Polishing an Amperometry Cell Working Electrode
These instructions are for conventional (nondisposable) working
electrodes only. Do not polish disposable electrodes.
When to Polish the Working Electrode
Polish carbohydrate gold, platinum, silver, and glassy carbon working
electrodes before initial installation in the amperometry cell. Do not
polish new AAA gold electrodes before installation.
After the working electrode is polished and installed, background
signal and analyte sensitivity require several hours to stabilize. Once
stabilized, do not polish the electrode unless you observe a loss of
signal or severe electrode recession.
Items Needed
Polishing kit (P/N 036313) shipped with the electrode. The kit
contains polishing pads (P/N 036321), a bottle of fine polishing
compound (P/N 036318), and a bottle of coarse polishing compound
(P/N 036319).
Pen that writes with indelible ink
Gloves
Tweezers
Working Electrode Polishing Procedure
NOTE To avoid electrode fouling, always wear gloves when
handling electrodes.
1. Follow the instructions in Section 9.24.2, Step 1 through Step 6 to
remove the working electrode from the cell.
2. Prepare the polishing pads:
a. Using indelible ink, label the plastic side of a pad to designate it
for use with the coarse polishing compound.
b. Label another pad for fine polishing compound; also designate
the working electrode type with which it will be used.