7.15 TC Periodic Maintenance
Clean the interior of the TC, using a lint-free cloth. Use paper to absorb any
liquid inside the module. The cleaner the module, the more effective the
sensors are at detecting excessive gas or humidity.
Check for excess condensation inside the TC. Condensation may build up
when the TC is operated in a highly humid environment and the majority of
applications are run at cold temperatures. To remove condensation, run the TC
at its maximum temperature for at least 72 hours.
Inspect the door seal for signs of damage. A defective seal impairs the
performance of the instrument. If the seal is defective, contact Dionex
Technical Support.
7.16 TC Annual Maintenance
Dionex recommends performing preventive maintenance annually. A DC/TC
Preventive Maintenance Kit (P/N 061796) is available.
Some IC columns should not be operated at temperatures above 70 °C
(158 °F). Remove all columns from the TC before beginning the
maintenance procedure above.