EasyManua.ls Logo

Dionex ICS-3000 - EG Specifications; Electrical; Environmental; Physical

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. 065031-04 1/08 339
EG Specifications
A.7 Electrical
A.8 Environmental
A.9 Physical
Main Power
Requirements
90 to 264 VAC, 47 to 63 Hz (Auto-sensing power supply; no
manual voltage or frequency adjustment required)
Typical input power: 18 W
Maximum line draw: 1 A at 110 VAC
Fuse
Requirements
Two IEC 60127-2 slow-blow fuses (P/N 954773) rated at 2 A,
250 V
Operating
Temperature
4 to 40 °C (40 to 104 °F)
Operating
Humidity
5% to 95% relative humidity, noncondensing
Operating Pressure Minimum: 14 MPa (2000 psi)
Maximum: 21 MPa (3000 psi)
Dimensions Height: 40.6 cm (16 in)
Width: 22 cm (8.75 in)
Depth: 55.9 cm (22 in)
Minimum required clearance behind the EG: 5 cm (2 in)
Weight 18 kg (40 lb)